The application of inductively-coupled plasma (ICP) spectrometry for
the analysis of solutions and slurries is reviewed. The critical
parameters which affect analytical performance are identified and
methods for the optimisation of the ICP for such applications
demonstrated.
Particle size of the suspended sample is shown to be the most
important factor in the analysis of slurry samples. It is demonstrated
that only particles less than 8 um in diameter reach the plasma even
when the maximal diameter injector tube (3 mm i.d.) is used.
Accordingly various methods for the reduction of particle size have
been investigated and milling shown to be the preferable approach. A
low cost method using blown zirconia spheres in a screw-cap plastic
bottle agitated on a flask shaker was developed. Using this method it
proved possible to analyse a wide variety of solid samples using slurry
atomisation using both ICP- atomic emission spectrometry (ICP-AES)
and ICP-mass spectrometry (ICP-MS) with aqueous calibration.
Results are presented for the elemental analysis of certified reference
material (CRM) soils, catalyst samples, a zeolite, titanium dioxide,
P.T.F.E. powder, and unfiltered natural waters by slurry atomisation
ICP-AES.
Comparison by other techniques showed excellent accuracy for major,
minor and trace elements and equivalent precision to solution analysis.
The CRM soils, catalysts and the zeolite were also analysed by ICP-MS
using slurry atomisation with similar success and enhanced accuracy at
the trace level. Aluminium proved to be an exception and generally
low recoveries (80-90%) were observed for this element using slurry
atomisation.
Finally the uses of the simplex optimisation algorithm to improve
analytical performance, particularly in ICP-AES, is discussed. A
computer program to facilitate the use of simplex optimisation for a
variety of instrumentation was written and its use in three different
analytical techniques demonstrated.
The necessary hardware and software modifications to a commercially
available ICP spectrometer to enable automated simplex optimisation
are described. The prospects for such intelligent self-optimising
instrumentation are discussed as is the potential extension of the
application of slurry analysis by ICP-AES and ICP-MS.
Date of Award | 1987 |
---|
Original language | English |
---|
Awarding Institution | |
---|
OPTIMISATION OF THE INDUCTIVELY-COUPLED PLASMA FOR THE ANALYSIS OF SOLUTIONS AND SLURRIES
NORMAN, P. (Author). 1987
Student thesis: PhD