Ultrasonic Doppler probe assessment by scanned hydrophone.

RC Chivers, J Adach, PR Filmore

Research output: Contribution to journalArticlepeer-review

Abstract

A method is described for using a scanned hydrophone to assess the transmit-receive response of ultrasonic Doppler transducers. The method was used on batches of six probes at each of the nominal resonant frequencies of 2, 4 and 8 MHz. The method has the advantage of showing unwanted field characteristics and identifying the element that may be causing them. The resonant frequencies and the overall sensitivities of the elements and probes are determined, the latter varying by as much as a factor of two.
Original languageEnglish
Pages (from-to)161-178
Number of pages0
JournalClin Phys Physiol Meas
Volume7
Issue number2
DOIs
Publication statusPublished - May 1986

Keywords

  • Evaluation Studies as Topic
  • Transducers
  • Ultrasonics

Fingerprint

Dive into the research topics of 'Ultrasonic Doppler probe assessment by scanned hydrophone.'. Together they form a unique fingerprint.