Optimising curve fitting techniques to look for standardisation of the analysis of defocus curves derived from multifocal intraocular lenses

  • Mid and South Essex NHS Foundation Trust
  • Southend Private Hospital
  • University of Plymouth
  • Plymouth Marjon University

Research output: Contribution to journalArticlepeer-review

37 Downloads (Pure)

Abstract

Introduction
To establish the most appropriate curve fitting method to allow accurate comparison of defocus curves derived from intraocular lenses (IOLs).

Methods
Defocus curves were plotted in five IOL groups (monofocal, extended depth of focus, refractive bifocal, diffractive bifocal and trifocal). Polynomial curves from 2nd to 11th order and cubic splines were fitted. Goodness of fit (GOF) was assessed using five methods: least squares, coefficient of determination (R2adj), Akaike information criteria (AIC), visual inspection and Snedecor and Cochran. Additional defocus steps at −2.25 D and −2.75 D were measured and compared to the calculated visual acuity (VA) values. Area under the defocus curve and range of focus were also compared.

Results
Goodness of fit demonstrated variable results, with more lenient methods such as R2adj leading to overfitting and conservative methods such as AIC resulting in underfitting. Furthermore, conservative methods diminished the inflection points resulting in an underestimation of VA. Polynomial of at least 8th order was required for comparison of area methods, but overfitted the EDoF and monofocal groups; the spline curve was consistent for all IOLs and methods.

Conclusions
This study demonstrates the inherent difficulty of selecting a single polynomial function. The R2 method can be used cautiously along with visual inspection to guard against overfitting. Spline curves are suitable for all IOLs, guarding against the issues of overfitting. Therefore, for analysis of the defocus profile of IOLs, the fitting of a spline curves is advocated and should be used wherever possible.
Original languageEnglish
Pages (from-to)887-896
JournalOphthalmic and Physiological Optics
Volume42
Issue number4
Early online date11 Apr 2022
DOIs
Publication statusPublished - 1 Jul 2022

Fingerprint

Dive into the research topics of 'Optimising curve fitting techniques to look for standardisation of the analysis of defocus curves derived from multifocal intraocular lenses'. Together they form a unique fingerprint.

Cite this