New real-time force monitoring device with enhanced accuracy for evaluation of applied force during the joint mobilization technique

  • Kau Pui Chia
  • , Ooi Ke Li
  • , Teh Shyong Yuong
  • , Om Prakash Singh
  • , Ahmad Athif Bin Mohd Faudzi
  • , Sundaram Sornambikai
  • , Jagannathan Madhanagopal*
  • *Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

Abstract

BACKGROUND: Force Monitoring Devices (FMDs) reported in the literature to monitor applied force during Joint Mobilization Technique (JMT) possess complex design/bulky which alters the execution of treatment, has poor accuracy and is unable to feel the resistance provided by soft tissues limits its usage in the clinical settings. OBJECTIVES: This study aims to develop a highly accurate, portable FMD and to demonstrate real-time monitoring of force applied by health professionals during JMT without altering its execution. METHODS: The FMD was constructed using the FlexiForce sensor, potential divider, ATmega 328 microcontroller, custom-written software, and liquid crystal display. The calibration, accuracy, and cyclic repeatability of the FMD were tested from 0 to 90 N applied load with a gold standard universal testing machine. For practical demonstration, the FMD was tested for monitoring applied force by a physiotherapist while performing Maitland's grade I to IV over the 6th cervical vertebra among 30 healthy subjects. RESULTS: The obtained Bland-Altman plot limits agreement for accuracy, and cyclic repeatability was -1.57 N to 1.22 N, and -1.26 N to 1.26 N, respectively with standard deviation and standard error of the mean values of 3.77% and 0.73% and 2.15% and 0.23%, respectively. The test-retest reliability of the FMD tested by the same researcher at an interval of one week showed an excellent intra-class correlation coefficient of r= 1.00. The obtained force readings for grade I to IV among 30 subjects ranged from 10.33 N to 45.24 N. CONCLUSIONS: Appreciable performance of the developed FMD suggested that it may be useful to monitor force applied by clinicians during JMT among neck pain subjects and is a useful educational tool for academicians to teach mobilization skills.

Original languageEnglish
Pages (from-to)829-836
Number of pages8
JournalTechnology and Health Care
Volume29
Issue number4
DOIs
Publication statusPublished - 2021
Externally publishedYes

ASJC Scopus subject areas

  • Biophysics
  • Bioengineering
  • Biomaterials
  • Information Systems
  • Biomedical Engineering
  • Health Informatics

Keywords

  • accuracy
  • Force monitoring device
  • joint mobilization
  • neck pain
  • reliability

Fingerprint

Dive into the research topics of 'New real-time force monitoring device with enhanced accuracy for evaluation of applied force during the joint mobilization technique'. Together they form a unique fingerprint.

Cite this