Evaluation of SARS-CoV-2 lateral flow device sensitivity: a comparison of systematic laboratory assessments with manufacturer-reported sensitivity

  • Public Health and Clinical Oversight team: Matthias E Futschik, Zareen Thorlu-Bangura, Sarah A Tunkel, Edward Blandford
  • Public Health and Clinical Oversight (PHCO), Clinical and Public Health Group, UK Health Security Agency, 10 South Colonade, Canary Wharf, London, E14 4PU, UK.
  • Science Group, UK Health Security Agency, London, UK.
  • Health Protection Operations (HP Ops) Group, UK Health Security Agency, London, UK.
  • Oxford University Hospitals NHS Foundation Trust
  • University of Oxford
  • National Institute for Health Research Health Protection Research Unit in Healthcare Associate Infections and Antimicrobial Resistance
  • Public Health Wales, Cardiff, UK.
  • Leeds Teaching Hospitals and University of Leeds
  • NHS England and NHS Improvement, London, UK.
  • Queen Mary University of London

Research output: Contribution to journalArticlepeer-review

Abstract

BACKGROUND: Self-testing for SARS-CoV-2 infection using lateral flow devices (LFDs) was a key component of the COVID-19 pandemic response; however, LFD performance has shown a high degree of variability. Between August 2020 and July 2023, the UK Health Security Agency (UKHSA), including predecessor organisations, undertook a three-phase SARS-CoV-2 test development and evaluation programme to independently evaluate commercially available SARS-CoV-2 LFDs, incorporating standardised laboratory assessment of test sensitivity. Here we describe results from a comparison of UKHSA laboratory assessment findings with manufacturer-reported LFD sensitivity data.

METHODS: The UKHSA assessed the sensitivity of LFDs, by laboratory testing of surplus clinical samples from a secondary healthcare setting. These data were compared with manufacturer-reported clinical sensitivity data and analytical sensitivity (limit of detection [LOD; 50% tissue culture infectious dose [TCID50]/mL]) from LFD instructions for use (IFU).

FINDINGS: UKHSA-determined LFD sensitivity ranged from 32 to 83%. Of 86 LFDs assessed, 73 included device sensitivity data in the manufacturers' IFU that claimed clinical sensitivity ≥85%, and 49 claimed clinical sensitivity ≥95%. No evidence of correlation was observed between manufacturer-reported test sensitivity and UKHSA determined test sensitivity, and no evidence of correlation was observed between manufacturer-reported test LOD and UKHSA-determined test sensitivity.

INTERPRETATION: Laboratory evaluation found no evidence of correlation between manufacturer-reported SARS-CoV-2 LFD sensitivity data and UKHSA laboratory-determined sensitivity, supporting previous reports of discrepancies. Our findings suggest that manufacturer-reported performance data and claims for SARS-CoV-2 LFDs should be interpreted with caution and support the need for independent monitoring and testing, and standardisation of analysis methodologies.

FUNDING: This study was funded by UK Department of Health and Social Care; UK Health Security Agency (formerly Public Health England and the National Health Service Test and Trace); and the University of Oxford NIHR Biomedical Research Centre.

Original languageEnglish
Pages (from-to)103454
JournaleClinicalMedicine
Volume87
DOIs
Publication statusPublished - Sept 2025
Externally publishedYes

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being
  2. SDG 16 - Peace, Justice and Strong Institutions
    SDG 16 Peace, Justice and Strong Institutions

Fingerprint

Dive into the research topics of 'Evaluation of SARS-CoV-2 lateral flow device sensitivity: a comparison of systematic laboratory assessments with manufacturer-reported sensitivity'. Together they form a unique fingerprint.

Cite this