TY - JOUR
T1 - Development and Demonstration of a Wireless Ultraviolet Sensing Network for Dose Monitoring and Operator Safety in Room Disinfection Applications
AU - Cullinan, Michael F.
AU - Scott, Robert
AU - Linogao, Joe
AU - Bradwell, Hannah
AU - Cooper, Leonie
AU - McGinn, Conor
PY - 2023/2/23
Y1 - 2023/2/23
N2 - The use of mobile ultraviolet-C (UV-C) disinfection devices for the decontamination of surfaces in hospitals and other settings has increased dramatically in recent years. The efficacy of these devices relies on the UV-C dose they deliver to surfaces. This dose is dependent on the room layout, the shadowing, the position of the UV-C source, lamp degradation, humidity and other factors, making it challenging to estimate. Furthermore, since UV-C exposure is regulated, personnel in the room must not be exposed to UV-C doses beyond occupational limits. We proposed a systematic method to monitor the UV-C dose administered to surfaces during a robotic disinfection procedure. This was achieved using a distributed network of wireless UV-C sensors that provide real-time measurements to a robotic platform and operator. These sensors were validated for their linearity and cosine response. To ensure operators could safely remain in the area, a wearable sensor was incorporated to monitor the UV-C exposure of an operator, and it provided an audible warning upon exposure and, if necessary, ceased the UV-C emission from the robot. Enhanced disinfection procedures could then be conducted as items in the room could be rearranged during the procedure to maximise the UV-C fluence delivered to otherwise inaccessible surfaces while allowing UVC disinfection to occur in parallel with traditional cleaning. The system was tested for the terminal disinfection of a hospital ward. During the procedure, the robot was manually positioned in the room by the operator repeatedly, who then used feedback from the sensors to ensure the desired UV-C dose was achieved while also conducting other cleaning tasks. An analysis verified the practicality of this disinfection methodology while highlighting factors which could affect its adoption.
AB - The use of mobile ultraviolet-C (UV-C) disinfection devices for the decontamination of surfaces in hospitals and other settings has increased dramatically in recent years. The efficacy of these devices relies on the UV-C dose they deliver to surfaces. This dose is dependent on the room layout, the shadowing, the position of the UV-C source, lamp degradation, humidity and other factors, making it challenging to estimate. Furthermore, since UV-C exposure is regulated, personnel in the room must not be exposed to UV-C doses beyond occupational limits. We proposed a systematic method to monitor the UV-C dose administered to surfaces during a robotic disinfection procedure. This was achieved using a distributed network of wireless UV-C sensors that provide real-time measurements to a robotic platform and operator. These sensors were validated for their linearity and cosine response. To ensure operators could safely remain in the area, a wearable sensor was incorporated to monitor the UV-C exposure of an operator, and it provided an audible warning upon exposure and, if necessary, ceased the UV-C emission from the robot. Enhanced disinfection procedures could then be conducted as items in the room could be rearranged during the procedure to maximise the UV-C fluence delivered to otherwise inaccessible surfaces while allowing UVC disinfection to occur in parallel with traditional cleaning. The system was tested for the terminal disinfection of a hospital ward. During the procedure, the robot was manually positioned in the room by the operator repeatedly, who then used feedback from the sensors to ensure the desired UV-C dose was achieved while also conducting other cleaning tasks. An analysis verified the practicality of this disinfection methodology while highlighting factors which could affect its adoption.
U2 - 10.3390/s23052493
DO - 10.3390/s23052493
M3 - Article
SN - 1424-8220
VL - 23
SP - 2493
EP - 2493
JO - Sensors
JF - Sensors
IS - 5
ER -